Device Specifications

Category

Specification

Sample Requirements

Type: Lithium Heparin Anticoagulant (whole blood, serum, plasma)

Volume: 90–120 uL

Bar Code: QR Code

Testing Parameters

Testing Time: 12 minutes/sample

Principle: Absorption spectroscopy, transmission turbidimetry

Method: Endpoint, kinetic, fixed time, turbidimetry

Temperature: 37 ± 0.3°C

Resolution: 0.001 Abs

Cross Contamination: None

Quality Control

QC & Calibration: Intelligent QC

Working Conditions

Temperature: 10–30°C

Humidity: 30%–70%

Technical Specifications

Light Source: 12V/20W halogen tungsten lamp, >2500 hours lifespan

Power Supply: AC 220V, 50Hz, 120VA

Optical System: Post-filter spectral, 8-wavelength synchronous detection (340, 405, 450, 505, 546, 600, 630, 850nm)

Display

Android 7.0 inch, 800x480, multi-point capacitive touch screen

Storage

500,000 results

Printer

Built-in thermal printer

Connectors

4 USB ports, 1 LAN port

Dimensions and Weight

Net Weight: 4.6 kg

Dimensions (LWH): 200 mm × 252 mm × 299 mm


Testing Parameters

UREA, Crea, GLU, ALT, AST, GGT, UA, HDL, LDL, TC, TB, DB, ALP, ALB, TG, AMY, CK, HBDH, LDH, CK-MB, CI, K, Na, Ca, PHOS, Mg, CRP, HbA1c, PT, APTT, FIB, TT, D-Dimer, ATIII, INR,